Quantitative Detection Kit (ELISA) for Rat Tumor Necrosis Factor Receptor Superfamily 14 (HVEM / TNFRSF14)

Instruction Manual of Quantitative Detection Kit (ELISA) for Rat Tumor Necrosis Factor Receptor Superfamily 14 (HVEM / TNFRSF14)

【Kit name】

Quantitative Detection Kit (ELISA) for Rat Tumor Necrosis Factor Receptor Superfamily 14 (HVEM / TNFRSF14)

【Use of the kit】

Quantitatively detect the content of tumor necrosis factor receptor superfamily member 14 (HVEM / TNFRSF14) in rat serum, plasma and related liquid samples.

【Detection principle】

This kit uses double-antibody two-step sandwich enzyme-linked immunosorbent assay (ELISA). Add the standard and the sample to be pre-coated to the rat tumor necrosis factor receptor superfamily member 14 (HVEM / TNFRSF14) antibody transparent enzyme label coated plate. After incubation for a sufficient time, wash to remove unbound components. Add enzyme working solution again, and after incubation for a sufficient time, wash to remove unbound components. Substrates A and B were added in sequence, and the substrate (TMB) was converted into a blue product catalyzed by horseradish peroxidase (HRP), which turned yellow under the action of acid. The concentration of factor receptor superfamily member 14 (HVEM / TNFRSF14) is positively correlated. The OD value is measured at a wavelength of 450 nm. Based on the OD value of the standard and the sample, the rat tumor necrosis factor receptor superfamily member 14 (HVEM / TNFRSF14) in the sample is calculated )content.

【Composition of the kit】

1

Enzyme coated plate

12 holes × 8

7

Developer A liquid

6mL

2

Standard product (120ng / L)

0.5mL

8

Developer B liquid

6mL

3

20 times concentrated washing liquid

25mL

9

Stop solution

6mL

4

Sample diluent

6mL

10

Instructions

1 serving

5

Standard dilution

6mL

11

Sealing film

2 sheets

6

Enzyme reagent

6mL

12

sealed bag

1

Remarks: The standard product is diluted with standard product dilution solution as follows: 120, 60, 30, 15, 7.5, 3.75 ng / L.

[Reagents and equipment needed but not provided]

1. 37 ℃ thermostat

2. Standard specification microplate reader

3. Precision pipettes and disposable tips

4. Distilled water

5. Disposable test tubes

6. Absorbent paper

【Steps】

1. Preparation: Remove the reagent kit from the refrigerator and re-equilibrate at room temperature for 30 minutes.

2. Mixing solution: dilute the 20-fold concentrated washing solution with distilled water to the original one.

3. Add standard products and samples to be tested: take a sufficient number of enzyme-coated plates and fix them to the frame. Set up standard wells, sample wells to be tested and blank control wells, record the positions of each well in the standard wells. Add 50μL of standard product; first add 10μL of sample to be tested, and then add 40μL of sample diluent (that is, the sample is diluted 5 times); blank control well is not added.

4. Incubation: Incubate in a 37 ° C water bath or thermostat for 30 minutes.

5. Wash the plate: discard the liquid, pat dry on the absorbent paper, fill each hole with the washing solution, let stand for 1min, shake off the washing solution, pat dry on the absorbent paper, repeat the washing 4 times (you can also use the washing machine to press Instructions for washing the board).

6. Add enzyme-labeled working solution: add 50μL of enzyme-labeled working solution to each well, without adding blank control wells.

7. Incubation: Incubate in a 37 ° C water bath or thermostat for 30 minutes.

8. Wash the plate: discard the liquid, pat dry on the absorbent paper, fill each hole with washing liquid, let stand for 1min, shake off the washing liquid, pat dry on the absorbent paper, repeat the plate washing 4 times (you can also use the washing machine to press Instructions for washing the board).

9. Color development: add 50μL of developer A solution to each well, and then add 50μL of developer B solution, mix for 30s with a plate mixer (or gently shake for 30s by hand), and avoid color development at 37 ℃ for 15min .

10. Termination: Remove the enzyme labeling plate and add 50μL of stop solution to each well to stop the reaction (the color changes from blue to yellow).

11. Determination: Zero the blank holes, and within 15 minutes after termination, measure the absorbance (OD value) of each well with a wavelength of 450 nm.

12. Calculation: According to the concentration of the standard product and the corresponding OD value, calculate the linear regression equation of the standard curve, and then calculate the corresponding sample concentration on the regression equation according to the OD value of the sample. You can also use various application software to Calculation. The final concentration is the actual measured concentration times the dilution factor.

[Sample requirements]

1. The sample cannot contain sodium azide (NaN3) because sodium azide (NaN3) is an inhibitor of horseradish peroxidase (HRP).

2. The specimen should be extracted as soon as possible after collection. The extraction should be carried out according to relevant literature. The experiment should be carried out as soon as possible after extraction. If the test cannot be performed immediately, the specimen can be stored at -20 ℃, but repeated freezing and thawing should be avoided.

3. The sample should be fully centrifuged, without hemolysis and particles.

【Precautions】

1. The experiment is carried out in strict accordance with the instructions, and the result of the experiment must be determined by the reading of the microplate reader.

2. If the enzyme-labeled coated board is not used up after opening, it should be put into a sealed bag and added with desiccant immediately.

3. It is recommended that all standards, samples and blank controls be tested in duplicate, and the average value is taken to reduce the experimental error.

4. Keep in mind that the sample has been diluted 5 times. The calculation result is multiplied by 5 to obtain the actual concentration of the sample.

5. If the color is too light, the substrate incubation time can be extended properly.

6. In order to avoid cross-contamination, the standard, sample and blank control should be replaced with a new tip every time one is added; the public components such as enzyme working solution, sample diluent and substrate should be cantilevered and not touch the microwell ; Do not reuse the sealing film.

7. The kit is used within the warranty period, and reagents of different batches should not be mixed.

8. Substrate B is sensitive to light and avoid prolonged exposure to light.

[Summary of operating procedures]

Prepare reagents, samples and standards


Add prepared samples and standards, and react at 37 ℃ for 30 minutes

Wash the plate 4 times, add the enzyme reagent, and react at 37 ℃ for 30 minutes

Wash the plate 4 times, add color developing solutions A and B, and develop at 37 ℃ for 15 minutes

Add stop solution

Read OD value within 15 minutes

Calculation

【examination range】

3.75ng / L-120ng / L

【specification】

96T / box

【Storage】

Store at 2-8 ℃, protected from light and moisture.

【Validity】

6 months

Banding Machine can use "paper tape" or "opp tape" to pack, and use hot-melt bonding to make the product packaging simple and beautiful. Suitable for paper products, medicines, cosmetics, electronic products and general light goods. It can also be matched with the conveyor belt according to the production needs to achieve unmanned packaging. To achieve the goal of saving manpower and improving packaging quality