Explain the severity of the air pressure test in the high and low temperature low pressure test chamber

The high and low temperature and low pressure test chamber should be able to maintain the air pressure in the following figure, with a tolerance of ± 5% or ± 0.1 kPa (whichever is greater), and a tolerance of ± 2 kPa when the severity level is 84 kPa .

Air pressure

Approximate altitude / m

kPa

mbar

1

10

31200

2

20

26600

4

40

22100

8

80

17600

15

150

13600

25

250

10400

40

400

7200

55

550

4850

70

700

3000

84

840

(See note 1, note 2)

Note 1: The standard atmospheric conditions from 86 to 106kPa cover altitudes below 1000m above sea level.

Note 2: The severity level of 84kPa is applicable to the tests that require samples at the lower limit of standard atmospheric conditions.

The test duration of the high and low temperature and low pressure test chamber should be preferentially selected: 5min, 30min, 1h, 4h, 16h.

Noodle Box

Noodle Box,Sealed Box For Pasta,Noodles Storage Case,Knife And Fork Storage Box

Taizhou Baiying Commdity Co., Ltd , https://www.baiyingplastics.com